Abstract
Background: Hemoglobin (Hb) A1c, a biochemical marker widely used in monitoring diabetes mellitus, can be quantitatively measured by various examining systems. However, significant errors still exist. In the present study, we evaluated the HbA1c level in five patients with compound heterozygotes by five different examining systems and our goal is to identify the existence of erroneous HbA1c measurement.
Methods: Blood samples collected from normal (no hemoglobin variants) and abnormal (compound heterozygotes) patients were analyzed by capillary electrophoresis technique and sequence analysis. The samples without HbA expression via above methods were further analyzed for HbA1c by ion exchange HPLC Variant II/ Variant II Turbo 2.0 (VII and VII-T 2.0), boronate affinity HPLC, capillary electrophoresis, and Tinaquant immunoassay.
Results: HbA1c expression were unexpectedly detected in the compound heterozygous samples by using additional examining systems: The HPLC VII and VII-T 2.0 detected HbA1c expression in two of five samples and failed to detect the abnormal HbA2 expression; the CE system detected HbA1c expression in one of five samples with abnormal HbA2 expression; the Ultra2 and PPI system detected the HbA1c expression of all samples without abnormal HbA2.
Conclusions: Five human samples without HbA expression were additionally detected with HbA1c expression with or without abnormal HbA2 expression by five analysis systems and the different examining assay potentially affected the test results. These results demonstrated that the limitations of current examining systems for monitoring patients with hemoglobin disorders highlighting the further improvement in the method of clinical HbA examination.
Introduction
Glycated hemoglobin (HbA1c), the glycated fraction of hemoglobin A, is a biochemical marker. The protein is formed via nonenzymatic glycation of the valine residue at the N-terminal of hemoglobin β-chain with glucose. HbA1c test is routinely utilized in monitoring long-term glycemic control and assessing the risk of complications [1,2]. In the 2010 guideline of the American Diabetes Association, HbA1c was recommended as one criterion for diabetes screening and diagnosis using a cut-off value of 6.5% (48 mmol/mol) [3]. Therapeutic strategies, according to HbA1c test, have been established as well [4]. Therefore, the accurate and precise measurement of HbA1c is extremely crucial to clinical practices. Currently, a variety of methods based on different principles are used for HbA1c measurement in clinical laboratories and these methodologies included cation exchange-high performance liquid chromatography (CE-HPLC), boronate affinity high-performance liquid chromatography (BAC), capillary electrophoresis (CE), and immunoassay. However, the results of HbA1c test from these methods may be influenced with patients’ pathophysiological conditions, such as hemolysis, reduced erythrocyte life span, technical interference of certain hemoglobin variants, or elevated HbF expression [5]. More recently, several studies, reported the measurement of HbA1c, were significantly affected by patients with Hb variants (HbAS, HbAE, HbAC, HbAD, HbAJ (Bangkok), HbAG (Taibei)) and increasing evidence suggested that the results of HbA1c in Hb variants detected by implementation of BAC method as well as Roche Tinaquant immunoassay were much less affected than the result of International Federation of Clinical Chemistry and Laboratory Medicine (IFCC) reference method [6–9].
Compound heterozygosity is one of the critical elements to cause genetic disease in human being. The measurement of HbA1c is closely associated with the screen or diagnosis of compound heterozygosity patients with Hb variants. Notably, few studies were reported in the HbA1c measurement from compound heterozygous patients without HbA expression by using these examining assays. In the present study, HbA1c in the blood samples collected from five compound heterozygous patients without HbA expression was analyzed by five common HbA1c detection systems to better understand the HbA1c measurement in clinical application.
Materials and methods
Samples
The study was approved by the Ethics Committee of Zhongshan Hospital of Sun Yat-sen University. A total of 40 whole blood samples with EDTA collected from previously tested specimens were conducted with HbA1c analysis by a Bio–Rad Variant II Turbo 2.0 analyzer. Those samples were also analyzed by capillary electrophoresis (Capillarys 2; Sebia, Lisses, France) to confirm the absence of Hb variants. HbA1c in the 40 samples without HA variants (4.4–14.4% HbA1c) were then analyzed with different assays. Furthermore, five blood samples without hemoglobin A expression obtained from previously tested clinical specimens underwent hemoglobin analysis by Sebia capillarys 2 system and the type of hemoglobin was confirmed by genotyping analysis. Five samples with compound heterozygotes were collected: the genotype of the first sample was αα/αα and βCD26/βCD41-42; the second sample was αα/αα and βIVS2-654/βNewYork; the third sample was similar to the second sample, αα/αα and βCD41-42/βNewYork; the fourth sample was αα/αα and βCD41-42/βJ-Bangkok; the fifth sample was -SEA/-α4.2-Q-Thailand and β/β. The above five patients were nondiabetic patients with normal fasting blood glucose. All blood samples were aliquot quadruplicate and frozen at −70°C before analysis.
Analysis methods
In present study, Bio–Rad Variant II system (Bio–Rad, U.S.A) was used as a comparative method for Hb1c measurement given, it gained National Glycohemoglobin Standardization Program Level Laboratory certification and was traceable to the Diabetes Control and Complications Trial Reference method. A total number of 40 normal samples (no Hb variants) and five samples from patients with compound heterozygotes were examined using five routine methods: (1) the HPLC Variant II system (Bio–Rad, U.S.A); (2) the HPLC Variant II Turbo 2.0 system (Bio–Rad, U.S.A); (3) the Ultra2 system (Trinity Biotech, U.S.A) using the BAC principle; (4) the Capillarys 2 Flex Piercing (C2FP) system (Sebia, France) using the CE principle; (5) the Roche Modular PPI system (Roche, Germany) using Tina-quant Hemoglobin A1c III principle. The mixed whole blood samples were adopted to transferring value-assignment in order to improve the comparability of those systems to Bio–Rad Variant II. To analyze the results detected from samples with five examining assays, Bland–Altman plots were performed. Ordinary linear regression was performed to regression and bias analysis. The percentage deviation plots were used to carry out bias estimation (−6.0–6.0%).
Statistical analysis
Statistical analyses were carried out using SPSS software version 17.0. The Student’s t-test was used to compare differences between two groups. Data were presented as means ± standard deviations (SDs). Statistically significant difference was defined as P value <0.05.
Results
HbA1c measurement in the patient without HA variants
The HbA1c values of 40 patients without HA variants were correlated with VII system detected by VII, VII-T 2.0, Ultra2, C2FP, and PPI system. The results were further analyzed with corresponding equations: y = 1.0514x − 0.2502 (R2 = 0.9969), y = 0.9991x − 0.1105 (R2 = 0.9965), y = 1.0178x − 0.1701 (R2 = 0.9957), and y = 0.9932x + 0.1404 (R2 = 0.9957). Bland–Altman plots (Figure 1) showed agreement between VII system and VII-T, Ultra2, C2FP, PPI systems and the 95% confidence interval (95% Cl) for the deviation between VII system and those systems was −7.5% to +7.5% (95% Cl VII-T: 0.3 to −0.59; Ultra2: 0.48 to −0.24; C2FP: 0.38 to −0.44; PPI: 0.49 to −0.31).
Bland–Altman difference plot applied to comparing HbA1c results of VII system with other systems
HbA1c measurement in the patients with compound heterozygotes
HbA1c is formed on the basis of HbA and the change of HbA can impact the formation of HbA1c. Blood samples without HbA are defined as no HbA1c expression according to the definition of the International Federation of Clinical Chemistry and Laboratory Medicine. Five blood samples with compound heterozygotes in our study were detected without HbA expression, indicating that there was no HbA1c expression. However, the five examining systems showed different results.
Two erroneous results for HbA1c were attained in the VII analyzer and the chromatograms of the first (Figure 2B), fourth (Figure 2E), and fifth (Figure 2F) samples were all labeled ‘Volts’, which meant abnormal electrophoresis signal and was considered unacceptable according to the manufactures’ instructions. However, the VII chromatograms for the second (Figure 2C) and third (Figure 3D) samples with Hb New York and β-thalassemia (Figure 2C,D) were not easily distinguished from a normal HbAA chromatogram (Figure 3A) and the proportions of the HbA1c (Table 1) and HbA0 (85.5% and 86.3% respectively) were produced. The peak of Hb NewYork0 was all mistakenly identified as HbA0.
The chromatogram of five double heterozygous carriers on Bio–Rad VⅡ analyzer
The chromatogram of five double heterozygous carriers on Bio–Rad VⅡ Turbo analyzer
Samples . | Genotypes . | Bio–Rad VII . | Bio–Rad VII-T . | Sebia C2FP . | Trinity Biotech Ultra2 . | Roche PPI . |
---|---|---|---|---|---|---|
1 | αα/αα; βCD26/βCD41-42 | Nr | Nr | Nr | 4.2% | 4.5% |
22 mmol/mol | 26 mmol/mol | |||||
2 | αα/αα; βIVS2-654/βNewYork | 4.3% | 4.5% | Nr | 4.1% | 4.2% |
23 mmol/mol | 26 mmol/mol | 21 mmol/mol | 22 mmol/mol | |||
3 | αα/αα; βCD41-42/βNewYork | 4.5% | 4.6% | Nr | 4.3% | 4.8% |
26 mmol/mol | 27 mmol/mol | 23 mmol/mol | 29 mmol/mol | |||
4 | αα/αα; βCD41-42/βJ-Bangkok | Nr | 4.7% | Nr | 4.7% | 3.8% |
28 mmol/mol | 28 mmol/mol | – | ||||
5 | -SEA/-α4.2-Q-Thailand; β/β | Nr | Nr | 3.9% | 5.3% | 5.7% |
– | 34 mmol/mol | 39 mmol/mol |
Samples . | Genotypes . | Bio–Rad VII . | Bio–Rad VII-T . | Sebia C2FP . | Trinity Biotech Ultra2 . | Roche PPI . |
---|---|---|---|---|---|---|
1 | αα/αα; βCD26/βCD41-42 | Nr | Nr | Nr | 4.2% | 4.5% |
22 mmol/mol | 26 mmol/mol | |||||
2 | αα/αα; βIVS2-654/βNewYork | 4.3% | 4.5% | Nr | 4.1% | 4.2% |
23 mmol/mol | 26 mmol/mol | 21 mmol/mol | 22 mmol/mol | |||
3 | αα/αα; βCD41-42/βNewYork | 4.5% | 4.6% | Nr | 4.3% | 4.8% |
26 mmol/mol | 27 mmol/mol | 23 mmol/mol | 29 mmol/mol | |||
4 | αα/αα; βCD41-42/βJ-Bangkok | Nr | 4.7% | Nr | 4.7% | 3.8% |
28 mmol/mol | 28 mmol/mol | – | ||||
5 | -SEA/-α4.2-Q-Thailand; β/β | Nr | Nr | 3.9% | 5.3% | 5.7% |
– | 34 mmol/mol | 39 mmol/mol |
Nr, no HbA1c value was reported for this sample by the system; ‘–’, no IFCC HbA1c value transferred.
Similarly, the chromatograms of the VII-T 2.0 analyzer for the first (Figure 3B) and fifth (Figure 3F) samples also did not show the results of HbA1c and also labeled ‘Volts’. The VII-T 2.0 chromatograms for the second (Figure 3C) and third samples (Figure 3D) also appeared normal with no indication of variants and the Hb New York appeared to elute as the HbA0 window. The VII-T 2.0 chromatogram for the fourth (Figure 3E) was different from the VII chromatogram and showed a HbA1c value (4.7% in Table 1). Checking the chromatogram, we found the proportion of P4 was very high (84.9%), which was different from normal chromatogram (under 1.5%). So, we assumed that the P4 window was the peak of Hb J-Bangkok.
Of the five heterozygotes analyzed by C2FP system, erroneous results were reported in only one sample (Figure 4F). The electropherograms for the first, second, third and fourth, (Figure 4B–E) all showed the ‘atypical profile’ flag due to identification of the abnormal electropherograms. In addition, four of five electropherograms misidentified the peak of Hb variants as the peak of Hb A0. For the fifth sample with Hb Q-H disease (Figure 4F), electropherogram of C2FP analyzer produced an erroneous ‘normal’ HbA1c (3.9% in Table 1) result and did not show the peak of HbH.
The chromatogram of five double heterozygous carriers on Sebia C2FP analyzer
All of those hemoglobin variants were undetectable on the HbA1c principle from PPI and primus Ultra2 systems. It was remarkable that Ultra2 and PPI systems all produced spuriously ‘normal’ HbA1c results for all five double heterozygotes (Table 1).
Discussion
The HbA1c value reflects the patient’s mean glycemic level in the past 6–8 weeks. Hemoglobinopathy alters the composition and structure of hemoglobin and may lead to misinterpretation of the HbA1c result. Hemoglobin variants have been reported to potentially affect the precision of current examining methods for HbA1c measurement [10–12]. However, HbA1c values measured in patients with compound heterozygotes are rarely reported.
HbA1c has been referred as one of major markers for diabetes diagnosis by the World Health Organization since 2011 [13]. However, HbA1c is still not a diagnostic criterion for diabetes in China yet. One of the reasons is that China is a population with a high prevalence of hemoglobinopathy and thalassemias [14–17]. The incidence of α- and β-thalassemias is 8.53% and 2.54% respectively [14]. Of note, the incidence of abnormal Hbs is 0.358% in Chaozhou city [15] and 0.4% in Dongguan city, China. The common α-globin variants are Hb Constant Spring, Hb Q-Thailand, and Hb G-Honolulu. The common β-globin variant was Hb E, Hb New York, Hb J-Bangkok, et al. [16]. Thalassemias combined with hemoglobin variants are a special physiopathological condition that cannot form HbA. Five cases in our article were double heterozygous samples that do not contain HbA and HbA1c. We tested those five samples by five examining systems in order to assess the specificity of common methods in the measurement of HbA1c.
HbA1c values can be evaluated by various methods based on the molecular charge (cation-exchange high-performance liquid chromatogram (CE-HPLC) and electrophoresis) or molecular structure (immunoassays, boronate affinity chromatography, and mass spectrometry). Both VII and VIIT2.0 belong to CE-HPLC assay, which can separate HbA1c from HbA because glycation of the N-terminal valine reduces the positive charge. In present five cases, the two principles reported two of five and three of five spurious HbA1c results respectively. Because Hb New York has the similar charge to HbA and can be eluted together with HbA Hb New York in the second and third cases, it was erroneously identified as HbA0 due to elution in the respective retention and Hb (New York)1c misidentified as HbA1c. Hb J-Bangkok is another common hemoglobin variant in China and its charge is different from HbA. Therefore, the chromatograms of VII and VIIT2.0 were shown abnormally. Lo et al. [18] reported that a case of spuriously normal HbA1c results due to misidentifying HbG Taipei as HbA0 by the variant II system. These results showed that CE-HPLC was obviously interfered by Hb variants with different charges and might result in erroneous HbA1c concentration.
C2FP is a new assay to evaluate HbA1c based on the separation of Hb property and charge by capillary electrophoresis. There is a strong consistency between the results of C2FP and VII [19]. Many literatures have reported that the resolution of C2FP is superior to CE-HPLC resulting from allowing the separation of many common and rare Hb variants from the HbA0 fraction [19–21]. Of five samples without HbA expression, C2FP system could detect HbA1c up to four samples. Although C2FP system misidentified Hb F as HbA0 in the first sample and Hb (New York) 0 as HbA0 in the second and third sample, the system did not display HbA1c values, which might prompt laboratories to pay more attention to those patients with hemoglobin fractions. However, in the fifth sample, C2FP system detected the HbA1c values without showing HbH expression and misidentified HbQ-Thailand as HbA0. Literatures had reported that C2FP system produced two inaccurate results of the 18 rare variants (Hb Silver Springs and Hb J-Broussais) [11]. For HbG Coushatta, C2FP system produced a HbA1c value different from the results detected by Tandem HPLC-capillary electrophoresis (IFCC reference method) [22]. Although C2FP system could separate many common and rare Hb variants from the HbA0 fraction, we also must focus on analyzing raw data of the different Hb variants and finding the problems of the electropherograms.
Ultra2 system using boronate affinity HPLC method is based on the fact that glycated and non-GHB are separated regardless of hemoglobin species and the Ultra2 system for Hb A1c seems to have less interference by Hb variants compared with the IFCC Reference Method [23]. PPI system is based on Tinaquant immunoassay using antibodies to target N-terminal glycated amino acids on the β chain for quantifying Hb A1c and the Hb A1c percentage is calculated according to the Hb A1c and Hb concentrations. Ultra2 system and Tinaquant immunoassay had been used as comparative methods in previous articles since these methods were less likely to show interference by Hb variants [11,24]. In our five cases, the Ultra2 system and PPI system measured total glycated hemoglobin and detected a normal HbA1c. But there was no Hb A in these samples and Hb A1c could not be detected. So, we assumed that the HbA1c measurement derived from patients without Hb A expression by those examining systems should not be used as an indicator of those people with average blood glucose level to screen and diagnose the diabetes mellitus.
Conclusions
Each examining system for HbA1c measurement could not eliminate the interference of double heterozygous in our work. The implementation of such methods may not generate trustworthy results for clinic application. Though C2FP system may be superior to other systems, it is important to know that the hemoglobinopathy can affect HbA1c measurement. Patients with compound heterozygous variants are suggested to use non-Hb-based method, such as fructosamine, glycated albumin or continuous glucose monitoring, to assess long-term glycemic control instead of Hb A1c measurement. Due to the high frequencies of hemoglobin variants, it is important to clarify these limitations when using these methods to measure Hb A1c.
Competing Interests
The authors declare that there are no competing interests associated with the manuscript.
Funding
This study was supported by Zhongshan science and technology bureau project [K2015B1002] and Zhongshan Health Bureau [2014J040].
Author Contribution
M.S. designed and performed the research, analyzed the data, and wrote the paper; D.W. guided and supervised the experiment process; W.W. detected the specimens; D.Z. detected the specimens; S.X. collected the specimens; X.W. detected the specimens; T.H. detected the specimens.